GEHC OEC 9800 ESP-15 C-Arm fluoroscopic X-ray system, GE Healthcare, Salt Lake City, UT.
FDA Recall Z-0990-2007 · Posted 2007-07-12
Advertisement
Recalling FirmGE OEC Medical Systems, Inc
Recall StatusTerminated
Product CodeJAA
Date Initiated2007-07-06
Date Terminated2013-02-01
Product Quantity4 units
Reason for Recall
The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
Firm Action
Consignees were notified by certified letter on 07/06/2007. They were told that a service rep would contact them to provide a label and an addendum to their operator's manual. A permanent software solution is under development and will be provided when it becomes available.
Root Cause
Other
Distribution Pattern
Worldwide, including USA, France, Singapore, Sweden, and Switzerland.