ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan

FDA Recall Z-0992-04 · Posted 2004-07-20

Recalling FirmR & D Systems,Inc
Recall StatusTerminated
Product CodeJPK
Date Initiated2004-04-15
Date Terminated2004-06-24
Product Quantity29494 vials of the normal controls and 15196 of the high controls

Reason for Recall

Clots in hematology controls resulted in low red blood cell counts when used in blood analyzers because control values were outside the expected range due to the clotting which took place after manufacturing.

Firm Action

Recall letters dated April 15, 2004 instructed consignees not to use the products to verify their analyzers and stated that the recalling firm is working on providing replacement products.

Root Cause

Other

Distribution Pattern

The products were shipped to California, France and Japan.