Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-35-45-RB-HFANL0-HC; 6) KCFW-12.0-35-45-RB-HFANL1-HC; 7) KCFW-9.0-35-38.5-RB-RTPS-100; 8) KCFW-10.0-35-RB-RTPS-100; 9) KCFW-9.0-38-55-RB-RAABE; 10) KCFW-9.0-38-70-RB-RAABE; 11) KCFW-9.0-38-80-RB; 12) KCFW-10.0-38-30-RB; 13) KCFW-10.0-38-40-RB; 14) KCFW-10.0-38-80-RB; 15) KCFW-10.0-38-RB; 16) KCFW-12.0-38-40-RB; 17) KCFW-12.0-38-80-RB; Flexor Introducers and Guiding Sheaths are designed to perform as a guiding sheath and/or introducer sheath. Flexor Introducers and Guiding Sheaths incorporate a Flexor shaft with a hemostasis valve and are provided with one or more dilators. These devices are available in various sizes, lengths and configurations. Configurations include differences in shaft construction, such as varying shaft stiffness and distal tip material and shape, dilator material, hydrophilic coating and/or distal radiopaque markers.
FDA Recall Z-0993-2026 · Posted 2025-12-23
Advertisement
Recalling FirmCook Incorporated
Recall StatusOpen, Classified
Product CodeDYB
Date Initiated2025-11-26
Product Quantity7952 units
Reason for Recall
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
Firm Action
Cook Medical issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/26/2025 by email and/or letter delivered via a courier. The notice explained the issue, risk to health, and requested the identification and return of the affected product. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred.
For questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Root Cause
Process control
Distribution Pattern
Worldwide distribution.