Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA, 5) TG90G01; b) InterStim X Clinician Therapy Application Software, Model Number A51300; and c) InterStim x "My Therapy" Patient Therapy Application Software, Model Number A52300

FDA Recall Z-0998-2022 · Posted 2022-04-22

Recall StatusTerminated
Product CodeEZW
Date Initiated2022-02-28
Date Terminated2026-09-30
Product Quantity1005 units

Reason for Recall

There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.

Firm Action

Medtronic distributed a correction notice to impacted U.S. implanting physicians, who had implanted patients with the Model 97800 InterStim X system prior to the release of the software update or purchased the TH90Q kit (in addition to notifying, Medtronic Sales Representatives and Medtronic Patient & Technical Services) on 28-Feb-2022. Medtronic distributed a correction notice to impacted U.S. hospital accounts, who have purchased the TH90Q kit containing the affected HH90 handset prior to the software updated release, on 04-Mar-2022, notifying consignees of this issue and the software update availability. Medtronic will also notify those consignees with the TH90PFA and TH90GFA kits where the affected software was installed.

Root Cause

Software Design Change

Distribution Pattern

US Nationwide distribution in the states of FL, IL, WI, TX, CA, ND, NC, KY, AR, MO, MN, OH, CO, KS, MS, SC, LA, WA, PA, IN, SD, MI, WV, AL, GA, TN, VA, IA.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑