AXIOM Luminos TF, AXIOM Iconos R100, AXIOM Iconos R200 and AXIOM Luminos dRF
FDA Recall Z-1000-2023 · Posted 2023-01-20
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2022-12-09
Product Quantity654 units
Reason for Recall
There is potential for the footrest to detach from the patient table during use.
Firm Action
A Customer Safety Advisory Notification (CSAN) was distributed to affected consignees beginning December 9, 2022. The CSAN was distributed to inform the customers about a preventative measure to exclude any potential risk of patient injury caused by the footrest detaching from the patient table during examination. Together with the CSAN, an addendum to the operator manual is distributed containing additional details and illustrations on how to securely attach the footrest to further improve the understanding of handling the footrest. Refer to the CSAN with addendum and copies of the Instructions for Use for each reported system.
Root Cause
Process control
Distribution Pattern
Worldwide distribution