Terminated

MagNA Pure LC Instrument; Catalog number 2236931.

FDA Recall Z-1001-03 · Posted 2003-07-15

Recall StatusTerminated
Date Initiated2003-05-27
Date Terminated2003-11-18
Product Quantity17

Reason for Recall

Potential for false negative patient results with software version 3.0

Firm Action

An Urgent Product Correction letter dated 5/27/03 was sent to each customer.

Root Cause

Other

Distribution Pattern

California, Georgia, Illinois, Maine, Michigan, Mississippi, New York Ohio, Oregon, Pennsylvania, Texas, Utah, Wisconsin and Canada.