Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Luminos Agile Max, Luminos dRF Max, and LUMINOS Lotus Max

FDA Recall Z-1002-2023 · Posted 2023-01-20

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2022-12-09
Product Quantity886 units

Reason for Recall

There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA). The system may be used on pediatric, adult, and bariatric patients. These devices are not for mammography examinations.

Firm Action

A Customer Safety Advisory Notification (CSAN) was distributed to affected consignees beginning December 9, 2022. The CSAN was distributed to inform the customers about a preventative measure to exclude any potential risk of patient injury caused by the footrest detaching from the patient table during examination. Together with the CSAN, an addendum to the operator manual is distributed containing additional details and illustrations on how to securely attach the footrest to further improve the understanding of handling the footrest. Refer to the CSAN with addendum and copies of the Instructions for Use for each reported system.

Root Cause

Process control

Distribution Pattern

Worldwide distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑