SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
FDA Recall Z-1003-2022 · Posted None
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeOVB
Date Initiated2022-02-24
Product Quantity23,368 units
Reason for Recall
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
Firm Action
An URGENT MEDICAL DEVICE PRODUCT ADVISORY dated 2/24/22 was sent to customers.
Inadvertent placement of SpaceOAR gel into a blood vessel and subsequent migration of the hydrogel outside of the pelvis, led to embolism. Boston Scientific will incorporate best practices regarding hydrogel placement during all SpaceOAR procedures into the IFU. The updated IFUs will be packaged with SpaceOAR and SpaceOAR Vue Systems after regulatory approvals are obtained for the IFU updates.
This Product Advisory contains the recommended IFU updates, located in Appendix 1, intended to:
-Raise awareness of embolism as an existing adverse event.
-Add new Warnings and Precautions that describe technique recommendations for the
placement of SpaceOAR and SpaceOAR Vue to reduce the unintentional placement of hydrogel.
- Add procedural instructions to reduce the unintentional placement of hydrogel in a blood vessel and steps to verify placement of the hydrogel following implantation.
NO product is being recalled and you are NOT required to return product to Boston Scientific. As per the IFU, SpaceOAR and SpaceOAR Vue System procedures should only be performed by physicians who have received appropriate training for proper hydrogel spacer deployment technique.
The Acknowledgement Form must be completed and returned even if you do not have any affected units. Return the Acknowledgement Form by Email: [email protected] or Fax to: Field Action Center 1-763-415-7708
Root Cause
Labeling design
Distribution Pattern
Worldwide - US Nationwide distribution.