Reflection, Microstable Acetabular Liner, REF 71740250, 32mm ID, 50-52 OD, 1 each, sterile EO. Component is a liner used to hold the femoral head used in total hip replacement arthroplasty procedures.
FDA Recall Z-1004-2009 · Posted 2009-04-21
Advertisement
Recalling FirmSmith & Nephew Inc
Recall StatusTerminated
Product CodeLPH
Date Initiated2008-11-21
Date Terminated2011-11-21
Product Quantity19 units
Reason for Recall
Liners measuring 28 mm ID and were mismarked as 32 mm.
Firm Action
All affected Smith & Nephew International Distributors were notified via e-mail dated November 24, 2008. In a letter titled "Urgent: Product Recall Notice - Immediate Action Required", distributors were instructed to immediately notify their accounts concerning the recall. All affected product is to be located and returned to the firm immediately.
Direct questions to Smith & Nephew, Inc. by calling 1-901-399-5441.
Root Cause
Process control
Distribution Pattern
International Distribution to Canada, France, Germany and Switzerland.