Vortex Surgical 25ga Illuminated Flex-Tip Laser Probe, Catalog Number VS0135.25;

FDA Recall Z-1004-2026 · Posted 2025-12-30

Recalling FirmVortex Surgical Inc.
Recall StatusOpen, Classified
Product CodeHQB
Date Initiated2025-11-19
Product Quantity221 units

Reason for Recall

Reason for the voluntary recall is some products in this lot are not passing through a 25ga cannula smoothly.

Firm Action

Vortex Surgical Inc. issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/19/2025 via email. The notice explained the issue, risk to health, and requested the use of the product be discontinued. The firm requested the response to be emailed to: [email protected] for replacement or refund. Customer Service will provide a return label for any unused products.

Root Cause

Component design/selection

Distribution Pattern

US and Japan

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.