Terminated

FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

FDA Recall Z-1005-2017 · Posted None

Recall StatusTerminated
Product CodeKPR
Date Initiated2016-11-18
Date Terminated2018-07-02
Product Quantity41

Reason for Recall

There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.

Firm Action

Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated December 2, 2016 to customers via USPS certified letter with returned receipt. The letter identified the affected product, problem and actions to be taken. Additionally, your local Philips representative has been provided details regarding this Device Correction and can be contacted if you need any further information or support concerning this issue: Technical Support Line: 1-800-722-9377

Root Cause

Component design/selection

Distribution Pattern

US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.