Terminated

Convenience kits containing regard IV Start Kits: (1) regard Clinical Packaging Solutions, Item Number: 830015013, LD00200M - Vag Delivery Mother; and (2) regard Clinical Packaging Solutions, Item Number: 830084003, LD00866C - Vag Delivery Mother.

FDA Recall Z-1005-2022 · Posted 2022-04-26

Recalling FirmROi CPS LLC
Recall StatusTerminated
Product CodeMLS
Date Initiated2022-03-18
Date Terminated2024-08-02
Product Quantity1,501 convenience kits

Reason for Recall

A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.

Firm Action

The recalling firm sent an email to the consignee on 3/18/2022 which said they have attached an Urgent Device Correction Notice for ROi CPS, LLC Convenience Kits that contain Regard branded IV Start Kits that were manufactured using PDI Prevantics products which were recalled by PDI. Attached to the email were the Urgent Device Correction notice dated 3/18/2022, a label template to be used to label the Regard IV start kits contained in the convenience kits, the original recall letter issued by PDI for their Prevantics products that were recalled, and a reply form. The Urgent Device Correction Notice dated 3/18/2022 explained the reason for the recall and listed the affected kit names and identification. The consignee was to review the inventory to determine if the affected kits were in inventory and to segregate and quarantine the kits noting that only the IV Start kit component within the kit was affected and all other items could be used. The customer was to use the attached label template to print warning labels to be affixed to the Regard IV Start Kits contained within the convenience kits. The label was to be applied to each IV Start Kit in a prominently visible location to end users of the kits and the location of the label should not cover any other critical product information found on the existing product label. The labels said 'URGENT PDI PREVANTICS DRUG RECALL DO NOT USE PREVANTICS SWABSTICK OR MAXI SWABSTICK IN THIS KIT PER THE PDI DRUG RECALL NOTICE DATED 2-4-2022 FORWARDED TO YOUR FACILITY." The consignee was to immediately notify any customers to whom the product was distributed or forwarded and the recall notification from the recalling firm, the notice from PDI, and the label template could be forwarded, as well. A reply form was enclosed for completion to indicate response to the notice. The customer was to respond regarding the number of kits in inventory and that they have read and understood the attached notice, forwarded t

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the state of Missouri.

Official Source

View FDA Recall →