Terminated

Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")

FDA Recall Z-1006-2009 · Posted 2009-03-09

Recall StatusTerminated
Product CodeMKJ
Date Initiated2008-12-15
Date Terminated2012-03-19
Product Quantity10, 735

Reason for Recall

Reliability issues - potential to shock a non shockable rhythm or not shocking a shockable rhythm;

Firm Action

Welch Allyn sent a Customer Recall Notice dated February 26, 2009, to all customers (customer list supplied by distributors). Welch Allyn will upgrade all units with software version 2.02 and lower to version 2.06.02 by Proc. AED10 Prod Service Min HW/SW Update Rev. E 820-0178-00.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution.