Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")
FDA Recall Z-1006-2009 · Posted 2009-03-09
Advertisement
Recalling FirmWelch Allyn Protocol, Inc
Recall StatusTerminated
Product CodeMKJ
Date Initiated2008-12-15
Date Terminated2012-03-19
Product Quantity10, 735
Reason for Recall
Reliability issues - potential to shock a non shockable rhythm or not shocking a shockable rhythm;
Firm Action
Welch Allyn sent a Customer Recall Notice dated February 26, 2009, to all customers (customer list supplied by distributors). Welch Allyn will upgrade all units with software version 2.02 and lower to version 2.06.02 by Proc. AED10 Prod Service Min HW/SW Update Rev. E 820-0178-00.
Root Cause
Component design/selection
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.