ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

FDA Recall Z-1007-2024 · Posted 2024-02-07

Recalling FirmGlobus Medical, Inc.
Recall StatusOpen, Classified
Product CodeOLO
Date Initiated2023-12-15
Product Quantity570

Reason for Recall

Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.

Firm Action

On December 15, 2023, the firm began to notify customers by phone. On January 9, 2024, firm began following up with letters. Customers were instructed to immediately return affected cannulas to Globus Medical and complete the letter's Response Card. For recall-related questions, please contact Globus Medical Complaint Handling Unit at [email protected]. Fax: 610-300-1342.

Root Cause

Device Design

Distribution Pattern

Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.