BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
FDA Recall Z-1008-2026 · Posted 2025-12-31
Advertisement
Recalling FirmBeaver-Visitec International, Inc.
Recall StatusOpen, Classified
Product CodeHQF
Date Initiated2025-12-05
Product Quantity5 units
Reason for Recall
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Firm Action
Beaver-Visitec International (BVI) notified consignees via emailed letter on 12/05/2025. Consignees were informed that the affected units can still be used. A BVI service technician will contact each consignee to schedule a visit. During this visit, the missing ferrites will be installed to restore the equipment to its fully compliant and validated configuration. BVI aims to complete the correction by December 31st, 2025.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.