Cardioplegia Pack BCD Vanguard Ratio 4:1 with Shunt
FDA Recall Z-1009-04 · Posted 2004-07-20
Advertisement
Recalling FirmCobe Cardiovascular, Inc
Recall StatusTerminated
Product CodeDTN
Date Initiated2004-05-25
Date Terminated2005-03-31
Product Quantity216 sets
Reason for Recall
Some Cardioplegia Delivery Sets, 4:1 w/shunt, were misassembled with 1:1 tubing sub-assemblies. This would provide 2 1/2x the desired cardioplegia.
Firm Action
Customers were notified by telephone on 5/25/2004 and by letter on 6/3/2004.
Root Cause
Other
Distribution Pattern
AL, AZ, FL, KS, MN, MO, NV, TX, WI. Foreign distribution to Canada.