fastener, fixation, nondegradable, soft tissue
FDA Recall Z-1009-2018 · Posted None
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeMBI
Date Initiated2017-10-25
Date Terminated2020-05-12
Product Quantity429 units
Reason for Recall
It was identified that the product was not assembled correctly causing the implant shaft to shift and fall out during the opening of the device, which potentially could break the device.
Firm Action
In October of 2017 an URGENT MEDICAL DEVICE RECALL LETTER was issued to customers requesting that any remaining product be quarantined and returned to recalling firm. Questions or concerns can be directed to 574-371-3071.
Root Cause
Process control
Distribution Pattern
TX, NC, FL, PA, IN, WI, VA, NY, NJ, AL, CT
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.