Open, Classified

Accula SARS-CoV-2 Test, REF: COV4100

FDA Recall Z-1009-2022 · Posted None

Recalling FirmMESA BIOTECH, INC
Recall StatusOpen, Classified
Product CodeQJR
Date Initiated2022-04-06
Product Quantity6162

Reason for Recall

SARS-CoV-2 Test has an increased potential for false positive results, which may lead to a delay in correct diagnosis and treatment.

Firm Action

On 04/06/22, recall notices were mailed to customers. Customers were informed that the reliability of negative test results is not affected. Customers with affected devices were asked to do the following: 1. Discontinue use of any affected product immediately 2. All affected product must either be returned or discarded. Replacement product will be provided for all returned and discarded product upon receipt of Customer/Distributor Response Sheet. 3. Notify all impacted users in your facility. If you have shipped any of these lots outside of your facility, you must notify those customers or facilities of this voluntary recall. 4. Complete and return the Customer/Distributor Response Sheet, which can be emailed to [email protected] Customers with questions are encouraged to contact the recalling firm at: +1-858-800-4929, or 1-800-955-6288, option 2.

Root Cause

Material/Component Contamination

Distribution Pattern

US Nationwide distribution in the states of CA, TX, MA, NY, FL, NC, NJ, GA, CO, HI, PA, WA, NY, VA, MD, KY, WI, IL, PR, ND, MI, TN, AZ, UT.

Official Source

View FDA Recall →