PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; Pulse Spray 3Fx135x20, UPN: H787125004155, REF: 12500415; and PULSE SPRAY 5F/45/10, UPN: H787125004235, REF: 12500423.
FDA Recall Z-1010-2024 · Posted 2024-02-07
Advertisement
Recalling FirmAngiodynamics, Inc.
Recall StatusOpen, Classified
Product CodeKRA
Date Initiated2023-12-18
Product Quantity7 eaches
Reason for Recall
Rotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.
Firm Action
An Urgent Medical Voluntary Medical Device notices dated December 18, 2023 was issued to distributors and end users with instructions to immediately stop using affected product, remove said product from inventory and return to AngioDynamics, Inc. The notice should be forwarded to all site affected product may have been distributed to and recipients are asked to complete and return the Reply Verification Tracking Form. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product.
Root Cause
Process control
Distribution Pattern
United States (MD, LA & UT), Austria, Canada and Spain.