Recall StatusTerminated
Date Initiated2004-05-24
Date Terminated2007-12-11
Product Quantity8,000 units
Reason for Recall
A software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the volume is made at the Verify I-DRAIN:XXXXML prompt.
Firm Action
Urgent Device Correction letters dated 5/24/04 were mailed to the peritoneal dialysis nurse of facilities identified as having patients who use the HomeChoice machines. The letters informed them of the potential overfill problem and requested that they review the patient list provided with the letter and verify that it was complete. The nurses were requested to communicate the information to their home patients, and visit all of their patients who use the device in the low I-Drain Alarm Volume mode to make an interim correction, permanently changing the I-Drain Alarm Volume from OFF to 0ML in the Make Adjustments menu. Permanent software corrections will be installed when the revised software becomes available.
Follow-up letters dated 1/12/05 were sent to the Home Care Coordinators, providing them with an addendum to the HomeChoice Trainer's guide, repeating the information from the 5/24/04 letter and providing additional information to safely disconnect from the HomeChoice system during therapy. The letters also included copies of the 1/14/05 letter sent directly to the HomeChoice home patients, as well as addenda to the Patient Help Cards and Patient At-Home Guide.
Distribution Pattern
Nationwide, Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, India, Ireland, Italy, Korea, Malaysia, Mexico, Nicaragua, the Netherlands, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uruguay, Venezuela and the United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.