Terminated

WAVES WCM Kidney Cassette Module

FDA Recall Z-1013-2022 · Posted None

Recall StatusTerminated
Product CodeKDN
Date Initiated2022-03-21
Date Terminated2024-08-16
Product Quantity75 cases (6ct)

Reason for Recall

There is a potential for the WCM cassette pump head leaking solution.

Firm Action

WMS communicated the recall through a recall letter on 03/15/2022. The letter instructed all customers to return the product to WMS for evaluation. The goal is 100% product location and recovery. In addition, the letter instructed all customers to fill out information on the recall notification and return to WMS. For like product not involved in the recall, a warning letter was issued. For any product not included in the recall but found to be deficient, customers will be instructed to contact WMS for replacement.

Root Cause

Process control

Distribution Pattern

WCM cassettes have been distributed to one domestic customer with two locations in Tennessee and one international distributor located in France.