B. Braun Infusomat Space Pump - US Version Infusion pump. Model number 8713050U.
FDA Recall Z-1015-2012 · Posted 2012-02-29
Advertisement
Recalling FirmB. Braun Medical, Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2011-07-12
Date Terminated2014-11-03
Product Quantity5769
Reason for Recall
If the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed, the clip catch on the inside of the pump door may break. If the clip catch is broken and the door opened, free flow protection is ensured. If IV set is then removed from pump without closing the roller clamp, the clip may not re-engage the tubing, leaving an open fluid path with free flow pote
Firm Action
B Braun sent a letter on August 16, 2011, to their customers, that detailed the proposed G03 implementation in response to reported customer complaints. The software upgrade with the door jam was agreed upon and the eight hospitals were upgraded in November 2011.
Root Cause
Device Design
Distribution Pattern
Product was distributed to hospitals in CA and MO.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.