Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number: AS4010B; The product is manufactured by: BioGenex Laboratories, Inc. 4600 Norris Canyon Road San Ramon, CA 94583 and Distributed by: Abbott Molecular, 1300 E. Touhy Ave - 2E, Des Plaines, IL 60018 Xmatrx Automated Staining System is intended for in vitro diagnostic use. It is suitable for fluorescent hybridization (FISH) applications.

FDA Recall Z-1016-2010 · Posted 2010-03-08

Recalling FirmBiogenex Laboratories
Recall StatusTerminated
Product CodeKEY
Date Initiated2009-01-23
Date Terminated2010-12-02
Product Quantity12 units

Reason for Recall

Under certain conditions, product system failure has the potential to cause operator injury due to splash of Hemo-De solution.

Firm Action

On 1/23/09, the firm issued an Advisory Notice to all its direct consignees, informing them of the affected product and providing instructions on the recall. Please contact BioGenex Technical Support at 1-800-421-4149 for any assistance you may need.

Root Cause

Device Design

Distribution Pattern

The firm distributed the product to one (1) distributor as a consignee of our product. The firm indicated that no customer has received the product directly from BioGenex, due to the fact that the product is exclusive distributed by Abbott Molecular.