Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

FDA Recall Z-1016-2023 · Posted 2023-01-25

Recall StatusOpen, Classified
Product CodeCEW
Date Initiated2022-12-06
Product Quantity342 kits

Reason for Recall

Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.

Firm Action

On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom