Terminated

Dimension Vista System BUN Flex Reagent Cartridge, Catalog number K1021, in vitro diagnostic.

FDA Recall Z-1017-2010 · Posted 2010-03-08

Recall StatusTerminated
Product CodeJJE
Date Initiated2009-06-16
Date Terminated2010-11-30
Product Quantity1254 cartons

Reason for Recall

Incorrect result: patient sample BUN results reported may be higher or lower than actual level.

Firm Action

Siemens issued an Urgent Field Safety Notice dated 6/16/09 to all affected customers instructing them to discontinue use of the Dimension Vista BUN Flex reagent cartridge lot 09061AA.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution: USA, Australia, Canada, Germany, Japan, and Korea.