Terminated

Medtronic Archer Super Stiff Guidewire 0.035 inch (0.89mm), Model ARCSJ200W. Intended to facilitate catheter placement and exchange during procedures in the aorta.

FDA Recall Z-1017-2013 · Posted 2013-03-28

Recall StatusTerminated
Product CodeDQX
Date Initiated2013-03-13
Date Terminated2013-11-21
Product Quantity225 units

Reason for Recall

Medtronic has identified an issue with a specific lot of the Archer 0.035 inch (0.89mm) Super Stiff Guidewire, model ARCSJ200W, where the PTFE (Teflon) coating has the potential to separate from the wire during in-vivo use.

Firm Action

Medtronic sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Medtronic requested any unused guidewires from the lot number to be quarantined, and returned by contacting the customer service phone number at 1-888-283-7868. A Customer Confirmation Certificate was requested to be faxed back to Medtronic at 651-367-0612.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Worldwide Distribution-USA including the states of FL, OH, and MN and the countries of Belgium, France Germany, Netherlands, Turkey, and United Kingdom.