Invisalign Express 10 REF 8512 Invisalign System Comprehensive REF 9000 Invisalign System Moderate REF 9001
FDA Recall Z-1017-2023 · Posted 2023-01-26
Advertisement
Recalling FirmAlign Technology Inc
Recall StatusCompleted
Product CodePNN
Date Initiated2022-12-05
Product Quantity14 cases (637 aligner and temples)
Reason for Recall
3D orthodontic planning software has a defect that leads to an issue where incorrect number of aligners may be produced (less or greater than the doctor requested) and incorrect packaging is provided.
Firm Action
On 12/05/2022, the firm initiated telephone calls with customer to inform them of a voluntary recall of selected Invisalign aligners. Starting on 12/09/2022, the firm sent an "URGENT Medical Device Correction" email to customers as a follow-up to a telephone call informing them that selected Invisalign aligners may have the potential clinical impact and therefore may result in unnecessary/unprescribed treatment due to software issue with the new Treatment Planning History feature within ClinCheck software.
Customers are instructed:
1. Cease/stop using and dispose of the affected Invisalign case(s), including the aligners and packaging, and treatment PDF form-printed or online
2. Notify patients to discard the impacted aligners, if they have already been provided to them. The impacted product can be discarded as per the Doctor Instructions for Use or Patient Use and Care Instructions.
As an immediate action, Align Technology has disabled the ClinCheck feature capability from the Treatment Plan History.
For questions - contact [email protected] or 888-822-5446.
Root Cause
Software design
Distribution Pattern
U.S. Nationwide distribution in the states of AZ, CA, CO, FL, KY, MA, NJ, NY, OH, OK, and WI.