Terminated

RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The Siemens 400 series system is intended for point of care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.

FDA Recall Z-1019-2010 · Posted 2010-03-08

Recall StatusTerminated
Product CodeCHL
Date Initiated2009-10-30
Date Terminated2015-09-04
Product Quantity998 units

Reason for Recall

Software: Reference sensor failure may result in inaccurate pH reuslts

Firm Action

An "Urgent Field Safety Notice" dated September 2009 was sent via overnight mail to customers and included software Version 3.5.2 with instructions for install that will detect and prevent the reporting of potentially errant test results. The customers are requested to please install the included software update immediately and return the attached Fax back form. Siemens will provide the mandatory update to all customers with the Rapidpoint 400 /405 test system. If you have any questions, please contact your local Technical Support at 508-359-3877.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution: USA and Australia, Austria, Bahamas, Barbados, Belgium, Bosnia Herzeg, Brazil, Canada, Canary Islands, Chile, Croatia, Czech Republic, Egypt, France, Germany, Holy See (Vatican City State) Hong Kong, Ireland, Italy, Japan, Macedonia, Malaysia, M¿xico, New Zealand Not assigned P.R. China, Poland, Portugal, Puerto Rico, Refurb.Systems U.S.A Republic Korea, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, U.A.E., United Kingdom