2.0 Mandible Locking Plate (MLP) Bender / Cutter (part number 329.143) Intended to be used for bending and/or cutting Mandible Locking Plates.
FDA Recall Z-1021-2015 · Posted 2015-01-28
Advertisement
Recalling FirmSynthes, Inc.
Recall StatusTerminated
Product CodeHXW
Date Initiated2014-12-15
Date Terminated2016-08-18
Product Quantity973
Reason for Recall
Certain lots have the potential for tip breakage while bending 2.0 mm thick plates. If device is used and breaks while in the Operating Room, surgical delay may occur while alternate bending or cutting instruments are located.
Firm Action
An Urgent Notice: Medical Device Recall, dated December 19, 2014, was sent to end users and sales consultants that described the product, problem, and action to be taken. End users/sales staff were requested to immediately stop use and remove affected product from inventory. Customers were asked to call and obtain a Return Authorization Number, and complete and return the Verification Section of the letter along with the devices to Synthes per the instructions provided. Questions should be directed to 610-719-5450 or your local DePuy Synthes sales consultant.
Root Cause
Finished device change control
Distribution Pattern
Worldwide Distribution -- US and Canada.