VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.
FDA Recall Z-1022-2015 · Posted 2015-01-28
Advertisement
Recalling FirmINNOKAS MEDICAL OY
Recall StatusTerminated
Product CodeMWI
Date Initiated2014-12-03
Date Terminated2015-04-24
Product Quantity20 units
Reason for Recall
If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.
Firm Action
A notification letter was sent to the distributors on 12/3/2014. The letter requests the devices be returned to the manufacturer for repair.
Root Cause
Software design
Distribution Pattern
Distributed in the US and Europe.