Model 511 Cardio Respiratory Monitor with Oximax technology with software version 5.0
FDA Recall Z-1023-04 · Posted 2004-07-20
Advertisement
Recalling FirmCAS Medical Systems, Inc.
Recall StatusTerminated
Product CodeBZQ
Date Initiated2004-05-24
Date Terminated2005-06-13
Product Quantity146 units
Reason for Recall
Device may not alarm for %SpO2 value limit violations if a specific feature is enabled
Firm Action
CAS Medical notified customers by telephone and followed with letter dated 5/24/04, advising users of the problem and to turn off the StatsSeconds feature.
Root Cause
Other
Distribution Pattern
CA, DC, IL, IN, GA, ME, MI, MN, NH, NJ, NY, PA, TX, WV. Foreign: Canada, Taiwan