MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
FDA Recall Z-1024-2026 · Posted 2026-01-13
Recalling FirmREPLIGEN CORPORATION
Recall StatusOpen, Classified
Date Initiated2025-12-10
Product Quantity33
Reason for Recall
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Firm Action
Repligen will repair the connections in each potentially affected MAVERICK device. In
the meantime, all customers have been instructed to refrain from using the device.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
U.S.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.