Depuy brand LCS Complete metal backed patella, porocoat, LRG; REF 1294-09-660.
FDA Recall Z-1026-05 · Posted 2005-07-21
Advertisement
Recalling FirmDepuy Orthopaedics, Inc.
Recall StatusTerminated
Product CodeNJL
Date Initiated2005-06-07
Date Terminated2006-05-02
Reason for Recall
There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
Firm Action
Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
Root Cause
Other
Distribution Pattern
Nationwide and international.