Trolley for Intera/NT MRI System
FDA Recall Z-1028-04 · Posted 2004-07-20
Advertisement
Recall StatusTerminated
Product CodeLNH
Date Initiated2004-05-13
Date Terminated2005-04-08
Product Quantity57 devices
Reason for Recall
Patient and MRI tabletop may fall if the tabletop is not positioned correctly on the MRI trolley.
Firm Action
The firm sent a letter via certified mail to all customers on May 13, 2004 instructing their customers to check the position of the tabletop to ensure it is correctly aligned on the trolley. The firm is implementing a design modification beginning May 13, 2004. A field service engineer will contact the customers to arrange for installation of the retrofit for the trolley.
Root Cause
Other
Distribution Pattern
Product was distributed to hospitals and medical centers nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.