Terminated

Circle C Dual Lumen Acute Hemodialysis/Apheresis Catheter Surgical Tray, DLC-600TEC, 12 Fr. x 15.2cm, Sterile, Manufactured by: Horizon Medical Products, Inc., One Horizon Way, P.O. Box 627, Manchester, GA 31816, 1-800-472-5221. The product is a clear rigid plastic surgical tray containing components required for implantation. Components of the inner tray are in a separate sterile pouch. The tray and components are then sealed in an outer clear plastic pouch.

FDA Recall Z-1029-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeKOC
Date Initiated2004-05-28
Date Terminated2005-01-12

Reason for Recall

The products have a potential defective seal that closes the product's OUTER POUCH. The surface of the INNER TRAY may no longer be sterile.

Firm Action

Consignees were notified via letter on 5/28/2004.

Root Cause

Other

Distribution Pattern

AL, CA, CO, FL, GA, IL, NJ, NY, OH, PA, SC, TN, TX & VA.