Computed Tomography, Models TSX-101A and TSX-021A.
FDA Recall Z-1031-04 · Posted 2004-07-20
Advertisement
Recalling FirmToshiba American Med Sys Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2004-03-17
Date Terminated2011-11-01
Product Quantity552
Reason for Recall
Erroneous patient data is entered. Previoius patient's demographics are recorded.
Firm Action
Firm issued Field Modification Instruction which describes the action to be taken by the firm''s own service technicians. Recall letters were sent to each site informing of the issue and provide caution regarding the problem prior to it''s getting fixed.
Root Cause
Software design
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.