Zimmer Silhouette Axial Torque Handle, Zimmer Spine, Inc., Minneapolis, MN; REF 7010-0151-01. Reusable instrument that is used to apply the appropriate torque to the lock nut that is part of the Silhouette Spinal Fixation System.

FDA Recall Z-1031-2009 · Posted 2009-03-04

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeHXC
Date Initiated2008-11-07
Date Terminated2009-12-22
Product Quantity76

Reason for Recall

Some of the torque wrenches apply less than the minimum torque to the screw locking cap when the torque indicators are aligned as described in the surgical technique.

Firm Action

The firm's sales force was notified by an Urgent: Device Recall letter dated 10/7/08, which instructed them to stop using the instrument and quarantine it immediately; physically count all affected product in their territory and record data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer Spine; and return the affected devices with the completed Inventory Return Certification Form. User accounts/implanting physicians were informed by an Urgent: Device Recall letter dated 12/10/08 and instructed to monitor their patients.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution --- USA including states of California, Minnesota, Ohio, Pennsylvania, and Tennessee, and countries of Denmark, Germany, Iran, Italy, Russia, Singapore, Spain, Switzerland, and United Kingdom.