CTL Medical Hex Driver MIS, 1/4" Square QC, 4.0 Ratcheting Ref 019.7050
FDA Recall Z-1031-2019 · Posted 2019-03-20
Advertisement
Recalling FirmCTL Medical Corporation
Recall StatusTerminated
Product CodeNKB
Date Initiated2018-07-05
Date Terminated2021-04-28
Product Quantity133 total devices
Reason for Recall
the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
Firm Action
The anti-toggle ring of the HEX Driver, 1/4", square QC, 40 can separate form the driver shaft.
Root Cause
Nonconforming Material/Component
Distribution Pattern