Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case
FDA Recall Z-1032-2023 · Posted 2023-01-27
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeLRO
Date Initiated2022-11-17
Product Quantity160 cases (480 units)
Reason for Recall
Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.
Firm Action
Medline Industries, LP notified consignees via email and first-class mail on 11/21/2022. Consignees were instructed to over-label the sterile kits with a warning label indicating Warning To Avoid Injury, Please Handle Blades with Care . For the Non-Sterile kits, they are to over label the affected kits indicating to remove the blades and replace from a different sterile supply. No product is to be returned. Customers were also instructed to complete and return the response form.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of AR, MD, WA, WI.