Terminated

Covidien Kangaroo Polyurethane Feeding Tube Radiopaque Line, Safe Enteral Connections 6.5 Fr/Ch (2.2 mm) x 36" (91 cm), Sterile; Catalog Number: 461438. Intended for pediatric feeding.

FDA Recall Z-1033-2016 · Posted 2016-03-02

Recalling FirmCovidien LLC
Recall StatusTerminated
Product CodeKNT
Date Initiated2016-01-21
Date Terminated2017-06-01
Product Quantity1,010 units

Reason for Recall

8FR 16 PVC tubes were packaged in a 6.5 FR 36 Polyurethane packs and 8FR 20 PVC tubes were packaged in an 8FR 20 Polyurethane packs

Firm Action

Medtronic/Covidien Customers notified via an Urgent Medical Device Recall letter, dated January 22, 2016. The letter identifies the affected product, instructs customers to discontinue use and quarantine any of the product, and return all affected inventory. Customers are also to complete the "Recalled Product Return Form." Questions or concerns can be directed to a Medtronic representative or Customer Service at 800-882-5878.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution -- US, to the state of CA; and, to Australia and Europe.