PULSAR MAX II (model nos. 1180, 1181 and 1280) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . PULSAR MAX II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. Two sensors are available; these adapt the pacing rate to the patient's changing metabolic demand.

FDA Recall Z-1035-05 · Posted 2005-07-28

Recalling FirmGuidant Corporation
Recall StatusTerminated
Product CodeLWP
Date Initiated2005-07-18
Date Terminated2007-01-05
Product QuantityINITIAL RECALL 361. EXPANDED RECALL 43.

Reason for Recall

A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.

Firm Action

INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.

Root Cause

Other

Distribution Pattern

worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.