Pointe Scientific Creatinine Reagent R2, Pointe Scientific, Inc., Canton, MI; Catalog 3-C7539-L, 7-C7539-R2-30, 7-C7539-R2-125, 7-C7539-R2-250, 8-C7539-R2-30, 8-C7539-R2-125 and 8-7539-R2-250. For the quantitative determination of creatinine in serum. For in-vitro diagnostic use only.

FDA Recall Z-1036-2010 · Posted 2010-03-09

Recall StatusTerminated
Product CodeCGX
Date Initiated2009-12-01
Date Terminated2010-10-15
Product Quantity625 vials

Reason for Recall

An additional 10% picric acid was added to the R2 component during production.

Firm Action

Consignees were notified by An Urgent: Medical Device Recall letter, dated 12/1/09, that identified the affected product. Customers were asked to immediately examine their inventory for the affected product and to notify any customers that the product may have been further distributed to by giving them a copy of the recall notice. Customers are to dispose of the affected product immediately. Questions should be directed to the Technical Support Department at (800)757-5313 or (734)487-8300. There is also a Recall Response form attached with this notice.

Root Cause

Process change control

Distribution Pattern

Worldwide Distribution -- USA, Bolivia, Greece, Guyana, India, Indonesia, Israel, Kenya, Korea, Malaysia, Mexico, Pakistan, Philippines, Russia, Serbia, Sri Lanka, Tanzania, Trinidad and Ukraine.