1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack
FDA Recall Z-1037-2022 · Posted None
Advertisement
Recalling FirmAmerican Contract Systems, Inc.
Recall StatusOpen, Classified
Product CodeOJH
Date Initiated2022-03-22
Product Quantity114 packs
Reason for Recall
Product was sterilized with a higher than specification EO concentration.
Firm Action
A customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following:
1) Review your inventory and segregate affected product. This product should not be shipped out to customers.
2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120.
3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness.
A Field Action Response Form should be completed and returned to [email protected], even if you do not have product inventory.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of IL, MO, NE, and TX.