Synchromed II infusion pump
FDA Recall Z-1040-04 · Posted 2004-07-20
Advertisement
Recalling FirmMedtronic Inc, Neurological & Spinal Division
Recall StatusTerminated
Product CodeFRN
Date Initiated2004-05-11
Date Terminated2005-11-23
Product Quantity14
Reason for Recall
The opposite version serial numbers were assigned to 20mL and 40mL SynchroMed II Model 8637 pumps. This size-type indication is used by the programmer to identify the pump size and can cause a limitation in the useable fill volume on a 40 mL pump and could cause some confusion related to the size of a 20 mL pump which identifies itself as a 40 mL pump.
Firm Action
The physicians have been notified via letter of the problem with the pump. Appointments will be made with the patients and a Medtronic representative to correct the implanted devices.
Root Cause
Other
Distribution Pattern
Within the United States and outside the United States.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.