Siemens Magnetom Harmony Mobile MRI System. Model Number: 4759309, Siemens Medical Solutions USA, Malvern PA, 19355
FDA Recall Z-1042-2008 · Posted 2008-03-25
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLNH
Date Initiated2007-12-01
Date Terminated2012-10-01
Product Quantity12 units
Reason for Recall
Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Firm Action
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution