Terminated

Siemens Magnetom Harmony syngo. Model Number: 7106714, Siemens Medical Solutions USA, Malvern PA, 19355

FDA Recall Z-1043-2008 · Posted 2008-03-25

Recall StatusTerminated
Product CodeLNH
Date Initiated2007-12-01
Date Terminated2012-10-01
Product Quantity7 units

Reason for Recall

Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.

Firm Action

The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution