Terminated

Latex-Free Symbiq Pump Set, convertible pin, 106 inch piggyback with backcheck valve, 2 CLAVE ports, distal microbore tubing and OPTION-LOK, MICRODRIP, non-DEHP 60 drops/mL; a sterile single patient use Rx device; Hospira Inc, Lake Forest, IL USA. made in Costa Rice; product number 160000401, list number 16000-01.

FDA Recall Z-1045-2010 · Posted 2010-03-09

Recalling FirmHospira Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2009-11-24
Date Terminated2012-01-29
Product Quantity45,537 sets

Reason for Recall

Company has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.

Firm Action

Hospira, Inc. issued an "Important Administration Set Information" letter dated November 24, 2009 to their direct accounts, informing them of the affected product. Consignees were instructed to return the defective product to the firm. For further information, contact Hospira Medical Communications at 1-800-615-0187.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- United States, Australia, Canada, Costa Rica, India, Mexico, Netherlands, Philippines, Singapore and United Arab Emirates.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.