Centurion PICC : CVI2650 ANGIO TRAY CVI3610 CENTRAL LINE INSERTION BUNDLE CVI4025 PICC INSERTION BUNDLE (P/S) CVI4190 CENTRAL LINE INSERTION BUNDLE CVI4190 CENTRAL LINE INSERTION BUNDLE CVI4330 PICC INSERTION TRAY CVI4340 MIDLINE INSERTION TRAY CVI4485 MOUNT SINAI CVP INSERTION KIT CVI515 VENOUS ACCESS KIT CVI650 INSERTION TRAY
FDA Recall Z-1045-2020 · Posted None
Advertisement
Recalling FirmCenturion Medical Products Corporation
Recall StatusTerminated
Product CodeOFF
Date Initiated2019-12-11
Date Terminated2020-09-04
Product Quantity29020
Reason for Recall
Incomplete seals on the sterile package may compromise the sterility
Firm Action
Centurion issued a notification emailed to Medline Industries,
Centurion's parent company. Centurion issued mailing letters via First Class Mail beginning on 12/12/2019, letter states reason for recall, health risk and action to take: Please identify all inventory that you have within your possession and segregate it in a location where it cannot be accessed for use. 2) If you have transferred possession of this product to another individual, to another department or to another location in your organization, you are required by FDA regulations to notify them of this recall
communication. Please ask that any affected product be returned to you for collection. 3) Affected product is not to be returned to Centurion, and should be destroyed/discarded. Document the destruction of all product on the enclosed accountability record. 4) Please complete the enclosed accountability record, whether or not you have product, and fax or email at
your earliest convenience to the attention of Lisa Carpenter, Director of Quality Assurance Centurion Medical Products Corporation, (517) 546-5400 Ext.1122.
Root Cause
Packaging process control
Distribution Pattern
Nationwide Foreign: GHANA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.