Terminated

Wallach Cryotip T-00337 Anorectal Tip, Teflon Coated, used with the Cryosurgical LL100 or LL50X Reference: 900220

FDA Recall Z-1046-2007 · Posted 2007-07-11

Recall StatusTerminated
Product CodeGEH
Date Initiated2007-06-06
Date Terminated2015-07-15
Product Quantity14 units

Reason for Recall

The tip may separate from the probe.

Firm Action

Wallach Surgical notified accounts by fax on 6/6/07. A follw-up telephone call will be made to the 70 domestic consignees. Foreign consignees notiifed by fax or email with follow-up telephone calls. Accounts are requested to return product and replace with an alternate product.

Root Cause

Device Design

Distribution Pattern

Worldwide, including USA, Canada, Australia, England, Greece, Jordan, Turkey, Hungary, Netherlands, Saudi Arabia, Spain, and Lithuania

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →