Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.

FDA Recall Z-1048-2013 · Posted 2013-04-01

Recall StatusTerminated
Product CodeJJE
Date Initiated2013-02-27
Date Terminated2014-01-14
Product Quantity2207 total

Reason for Recall

Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

Firm Action

Siemens initiated their recall of this product on February 27, 2013 by sending an Important Customer Notification letter to consignees by FedEx. The letter identified the product, problem, and actions to be taken by the customers.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution.